Calibration | 5-23
Set Cal Timers - Cal Gas Duration
This menu option allows you to define how long the RCU should apply
the span and zero calibration gases to the sensor during an automatic
calibration or verification. For a verification, the system uses the span gas
inlet for the verify gas.
To find out how long it takes for your calibration gases to reach the sen-
sor and stabilize, select the Inject Cal Gas menu option and record the gas
stabilization period. Then use this length of time as your calibration gas
duration (we suggest that you also add a buffer to this length of time to be
certain that the gas has stabilized).
1. Select
Cal Gas Duration
from the Set Cal Timers menu option.
You will be prompted to enter the new span gas duration time:
O2 Span Time = xx:yy
Enter new time: Mn:Sc
2. Enter the new span gas duration time, then press the
Enter
key. If
performing an automatic verification, enter the verify gas time here.
Mn equals minutes and
Sc
= seconds. As you type the new span gas
duration, it overwrites the Mn:Sc text. If you wish to use the currently
displayed time, press the
Enter
key without entering a new value.
At this point you will be prompted to enter the Zero gas duration time:
O2 Zero Time = xx:yy
Enter new time: Mn:Sc
3. Enter the new zero gas duration, then press the
Enter
key. (If perform-
ing automatic verifications, this time value is ignored since only a
single gas is used.) Mn equals minutes and Sc = seconds. As you type
the new zero gas duration time, it overwrites the Mn:Sc text. If you
wish to use the currently displayed time, press the
Enter
key without
entering any text.
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...