Calibration | 5-21
You will then be prompted to enter the time of day when the first calibra-
tion of the cycle should begin:
Cal time is YY:YY
Enter new time Hr: Mn
5. Enter the time you wish the first calibration or verification to begin,
then press the
Enter
key. The system uses military time, so to enter
4:00 PM., you would type 16:00. If, for example, you set the cycle
frequency at 8 hours, and specify the first calibration to take place at
4:00 PM., the next automatic calibration or verification will take place
at midnight.
Set Cal Timers - Set Time and Date
This menu option allows you to set the date or time for the internal con-
trol unit clock.
Make sure the current time and date are correct before setting any
auto cal/auto verify parameters.
1. Select
Set Time & Date
from the Set Cal Timers menu.
You are prompted to enter the new time:
The time is: YY:XX
Enter new time: HR: Mn
2. Enter the new time, then press the
Enter
key. As you type the new
time, it overwrites the Hr: Mn text. The Series 2000 control unit uses
military time so 16:00 equals 4:00 PM., and 04:00 equals 4:00 AM. If
the current time is correct, press the
Enter
key without changing the
value.
NOTE
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...