Installation
|
3-17
EMC Grounding, Shielding, and Noise Protection
For EMC purposes, under no circumstances should you leave cable
shields disconnected at one end or both ends of the cable (sensor or
control unit or other device).
You must use twisted-pair cable in rigid metal conduit or use twisted pair
cable with an overall braided shield. All cable shields or conduits connect-
ing to the control unit must be chassis grounded.
EMC Grounding Method
•
Shield Ring Method
Connect all shields for that conduit entry (other than power) to a sup-
plied shield terminal ring (see Figure 3-16). This shield ring is a stain-
less steel ring with a metal tab. Place the shield ring under the conduit
nut. Crimp the shields from all cables for that conduit entry to a 1/4”
female quick disconnect, then push it onto the tab that sticks out of
the conduit shield ring. Keep shields as short as possible.
•
Ground Stud Method
Connect all cable shields for that conduit entry to the grounding stud
closest to that conduit entry hole.
Figure 3-16.
Direct shield
grounding method.
GROUND STUD METhOD
ShIElD RING METhOD
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...