6-2 | Thermox CEM O2/TM Analyzer
System Messages
Auto Cal Pending
This message indicates that an automatic calibration is pending. Once the
current calibration or verification is complete, the automatic calibration
will begin. This message will only appear if the Auto Calibration option is
installed and enabled.
Auto Verify Pending
This message indicates that an automatic verification is pending. Once the
current calibration or verification is complete, the automatic verification
will begin. This message will only appear if the Auto Calibrate option is
installed and enabled.
Calibration Aborted
This message indicates that the calibration has been aborted by the user.
Calibration Required
This message is displayed to inform the operator that a regular calibra-
tion is required (after a primary calibration is performed you must al-
ways perform a regular calibration). During a primary calibration the cell
temperature is changed and you must wait for one hour before a system
calibration can be performed. The system calibration will determine the
constants needed for the Nernst equation.
If you see this message, select the Calibrate Key and perform a calibration.
This message will not clear from the display by removing or restoring
power to the control unit.
Cal Verify Aborted
This message notifies the operator that the verification was successfully
aborted by the system.
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...