Overview | 1-5
Startup Checklist
Review the Installation chapter (Chapter 3) and installation drawing
(Appendix D).
Install the sensor on the process. The ambient temperature must not
exceed 50°C (122°F).
Install the control unit. The ambient temperature must not exceed
50°C (122°F).
–
Do not create any additional conduit entries in the enclosure.
Interconnect the control unit and the sensor per the appropriate inter-
connecting drawing.
–
The specified cable must be used in order to insure proper opera-
tion of the systems.
–
Keep the interconnecting located in the control unit as short as
possible. A service loop need not be used.
–
Maintain the ordered wire pairs per the drawing.
–
Terminate the “shields” and/or “drains” of the cable properly.
Connect AC mains supply to the control unit and the sensor. Make
sure that the AC voltage is within specified limits.
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...