4-26 | Thermox CEM O2/TM
Alarm Key
Alarms 1 and 2 are hard wired for the watchdog timer and service
alarms. They are not selectable in software.
Oxygen alarms are disabled during a primary or regular calibration.
In addition, alarms are disabled during start-up until the sensor
reaches its proper operating temperature.
See the “Wiring” section in the Installation chapter for help on alarm con-
nections and how alarms are labeled on the control unit wiring card.
The Alarm Key allows you to set control unit alarm information. This
includes the following:
•
Define whether Alarm 3 activates based on oxygen levels, the start of a
calibration or the start of a verification. (Select Function).
NOTE
Figure 4-6.
Alarm Key menu.
Alarm Key
Alarm
Set Points
Alarm
Configure
Relay
Configure
Exception
Log
Select
Function
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...