Controller / User Interface | 4-19
Once you select what type of process readings the analog output
should track using the Select Function menu option, decide what
range of these readings the analog output range will cover (Set Cur-
rent Range).
Decide whether to enable output filtering, and if enabled, how much
to smooth out abrupt changes in readings (Output Filtering).
Analog Range Key menu options are presented in the order in which
they appear on your control unit screen, from top to bottom. Figure 4-5
provides an overview of the functions covered by each Analog Range Key
menu option.
Figure 4-5.
Overview of Analog
Range Key functions.
ANAlOG KEY
MENU OPTIONS
FUNCTIONS/OPTIONS
Set Current Range
Range of readings that the analog output represents
(for example: 4% O
2
= 4 mA; 10% O
2
= 20 mA).
Set Track/Hold
Hold output at last process reading during a cal or
verify
Set Current Mode
0 to 20 mA
4 to 20 mA
Output Filtering
Smooth out variations in readings
Select Function
O
2
Cell Temperature
Cell Millivolts
Thermocouple Millivots
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...