Maintenance and Troubleshooting | 6-3
Power Down Detected
The system detected that the power was removed. For example, a power
outage would cause this message to display. You can acknowledge this
system message by setting or reading any user-programmable value and
this will clear the message from the display. The message will not clear by
removing and restoring power to the unit.
Primary Calibration
This message appears when the system is performing a Primary Calibra-
tion. The message continues to be displayed during the one-hour recov-
ery period after the primary calibration.
System Calibrating
This message indicates that a system calibration is in progress.
System Verifying Cal
This message indicates that a system verification is in progress. It will ap-
pear during the entire automatic verification cycle or during the recovery
period for a remote or manual verification.
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...