4-22 | Thermox CEM O2/TM
where y is the analog output port and xx is the value that the current
output currently represents.
5. Type the value to correspond to the 4 (or 0) mA analog output.
Set Track/hold Option
This menu option allows you to decide what signals to send to the se-
lected analog output port during a calibration or verification. You have the
option to either hold the last process reading or track calibration readings.
Separate decisions can be made for verify and calibrate operations, so, for
example, you can elect to hold the last process reading during verifica-
tions, but track calibration gas readings during calibrations.
Setting track or hold
1. Select
Set Track/Hold
from the Analog Range Key after selecting an
analog output port.
2. You can choose to either
hold
process readings or
track
calibration
readings on the analog output port. Separate choices can be made for
calibration and verify operations.
For calibration operations, select one of the following:
Hold during cal
or
Track during cal
After you select one of these options, your choice will be confirmed on the
display:
Out # 1 will TRACK
during Calibration
or
Out # 1 will HOLD
during Calibration
For verify operations, select one of the following:
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...