Controller / User Interface | 4-21
Set Current Range Option
This menu option allows you to define the range of readings that the
0-to-20 or 4-to-20 mA (milliamp) output will represent. You can also
choose to set a reverse analog output range, so for example, 0 mA can cor-
respond to the high end of the oxygen range, and 20 mA can correspond
to the low end of the oxygen range.
The range of readings the current output represents is also a function
of what type of process readings the analog output port tracks (see the
“Select Function” section later in this chapter for the types of readings the
current output can represent).
Setting the current range
To set the range of readings that the selected current output represents:
1. Select
Set Current Range
from the Analog Range Key menu after
selecting an analog output port.
2. If tracking oxygen, select whether you want to enter the 20 mA value
using a percent or PPM range.
You will be prompted to define the new value to correspond to the 20
mA analog output:
Out # y 20 ma = xx
New Value?
where y is the analog output port, and xx is the present value that the
current output currently represents (for example, 10 % O2). Although
we use output port #1 for this example, the output port that will ap-
pear depends on the analog port you selected.
3. Type the value to correspond to the 20 mA analog output.
4. If tracking oxygen, select whether you want to enter the 0 or 4 mA
value using a percent or PPM range.
You will then be prompted to define the value to correspond to the 0
or 4 mA analog output:
Out # y 4 ma = xx
N e w V a l u e ?
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...