4-30 | Thermox CEM O2/TM
5. Select whether you want to enter the Alarm 4 set point using a percent
or PPM range.
You are then prompted to enter the Alarm 4 set point value:
Alarm #4 =xx
New Value?
6. Enter the Alarm 4 set point value.
7. Press the
Cancel
Key or the
Enter
Key to then exit this menu option.
Alarm Configure
This menu option allows you to define whether alarms should trigger
based on a high alarm or a low alarm condition.
Whether Alarm 3 is set high or low applies only if it is set to track
oxygen levels (% O
2
). If Alarm 3 is set to trigger at the start of a cali-
bration or verification, this option does not apply.
If you select Hydrocarbon Upset as the Sensor Config. option from the
Setup Key menu, Alarm 4 is fixed as both a high oxygen alarm and the
alarm that triggers during a hydrocarbon upset condition. In this case,
you will not be able to select Alarm 4 as a low oxygen alarm. Alarm 3,
however, can still be selected as a high or low oxygen alarm.
Configuring the alarms
1. Select
Alarm Configure
from the Alarm menu.
2. Define Alarm 3 and Alarm 4 as high or low alarms.
3. Press the
Cancel
Key to then exit this menu option.
NOTE
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...