5-16 | Thermox CEM O2/TM
Cal Gas Value
This menu option allows you to enter the zero calibration gas value to
match the calibration gas cylinder you use to calibrate your system. To set
your verify gas value, use the Configure Verify menu option instead.
• Span Gas - high calibration gas
• Zero Gas - low calibration gas
The Span Gas value you enter must be higher than the Zero Gas
value by at least one decade (for example, 1% and 10% O2).
Entering Cal Gas Values
1. Select
Cal Gas Value
from the Calibrate key menu.
2. Choose whether you wish to enter the span gas value using a percent
or PPM scale.
You are then prompted to enter the span gas value:
Span Gas = xx
New Value?
3. Enter the span gas value, then press the
Enter
key. If you wish to keep
the current zero gas value unchanged, press the
Enter
key without
entering a value.
4. Choose whether you want to enter the zero gas value using a percent
or PPM scale.
You are then prompted to enter the zero gas value:
Zero Gas = xx
New Value?
5. Enter the zero gas value, then press the
Enter
key. If you wish to keep
the current zero gas value unchanged, press the
Enter
key without
entering a value.
NOTE
Summary of Contents for Thermox CEM O2/TM
Page 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Page 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Page 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Page 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Page 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Page 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Page 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Page 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Page 137: ...Service and Parts 7 3 This page intentionally left blank...
Page 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Page 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Page 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Page 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Page 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...