Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
30
:WAVeform Commands
:WAVeform:POINts
(see
page 790
)
Command Syntax
:WAVeform:POINts <# points>
<# points> ::= {100 | 250 | 500 | 1000 | <points mode>}
if waveform points mode is NORMal
<# points> ::= {100 | 250 | 500 | 1000 | 2000 | 5000 | 10000 | 20000
| 50000 | 100000 | 200000 | 500000 | 1000000 | 2000000
| 4000000 | 8000000 | <points mode>}
if waveform points mode is MAXimum or RAW
<points mode> ::= {NORMal | MAXimum | RAW}
The :WAVeform:POINts command sets the number of waveform points to be
transferred with the :WAVeform:DATA? query. This value represents the
points contained in the waveform selected with the :WAVeform:SOURce
command.
For the analog or digital sources, the records that can be transferred
depend on the waveform points mode. The maximum number of points
returned for math (function) waveforms is determined by the NORMal
waveform points mode. See the :WAVeform:POINts:MODE command (see
page 642
) for more information.
Only data visible on the display will be returned.
When the :WAVeform:SOURce is the serial decode bus SBUS1, this
command is ignored, and all available serial decode bus data is returned.
Query Syntax
:WAVeform:POINts?
The :WAVeform:POINts query returns the number of waveform points to be
transferred when using the :WAVeform:DATA? query. Setting the points
mode will affect what data is transferred (see the :WAVeform:POINts:MODE
command (see
page 642
) for more information).
When the :WAVeform:SOURce is the serial decode bus SBUS1, this query
returns the number of messages that were decoded.
Return Format
<# points><NL>
<# points> ::= {100 | 250 | 500 | 1000 | <maximum # points>}
if waveform points mode is NORMal
<# points> ::= {100 | 250 | 500 | 1000 | 2000 | 5000 | 10000 | 20000
| 50000 | 100000 | 200000 | 500000 | 1000000 | 2000000
| 4000000 | 8000000 | <maximum # points>}
if waveform points mode is MAXimum or RAW
N O T E
The <points_mode> option is deprecated. Use the :WAVeform:POINts:MODE command
instead.
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...