:SBUS<n> Commands
25
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
467
:SBUS<n>:CAN:TRIGger:PATTern:ID
(see
page 790
)
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:ID <string>
<string> ::= "nn...n" where n ::= {0 | 1 | X | $}
<string ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $}
The :SBUS<n>:CAN:TRIGger:PATTern:ID command defines the CAN
identifier pattern resource according to the string parameter. This pattern,
along with the identifier mode (set by the
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE command), control the identifier
pattern searched for in each CAN message.
If the string parameter starts with "0x", it is a hexadecimal string made up
of hexadecimal and X (don't care) characters; otherwise, it is a binary
string made up of 0, 1, and X (don't care) characters.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:ID?
The :SBUS<n>:CAN:TRIGger:PATTern:ID? query returns the current settings
of the specified CAN identifier pattern resource in the 29- bit binary string
format.
Return Format
<string><NL> in 29-bit binary string format
Errors
•
"- 241, Hardware missing"
on page 753
See Also
•
"Introduction to :TRIGger Commands"
on page 589
•
":SBUS<n>:CAN:TRIGger:PATTern:ID:MODE"
on page 468
•
":SBUS<n>:CAN:TRIGger:PATTern:DATA"
on page 465
N O T E
The ID pattern resource string is always 29 bits. Only 11 of these bits are used when the
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE is STANdard.
A string longer than 29 bits is truncated to 29 bits when setting the ID pattern resource.
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...