Root (:) Commands
6
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
173
:OVLRegister (Overload Event Register)
(see
page 790
)
Query Syntax
:OVLRegister?
The :OVLRegister query returns the overload protection value stored in the
Overload Event Register (OVLR). If an overvoltage is sensed on a 50
Ω
input, the input will automatically switch to 1 M
Ω
input impedance. A "1"
indicates an overload has occurred.
Return Format
<value><NL>
N O T E
You can set analog channel input impedance to 50
Ω
on the 300 MHz, 500 MHz, and 1 GHz
bandwidth oscilloscope models. On these same bandwidth models, if there are only two
analog channels, you can also set external trigger input impedance to 50
Ω
.
Table 56 Overload Event Register (OVLR)
Bit
Description
When Set (1 = High = True), Indicates:
15-10
---
(Not used.)
9
Channel 4 Fault
Fault has occurred on Channel 4 input.
8
Channel 3 Fault
Fault has occurred on Channel 3 input.
7
Channel 2 Fault
Fault has occurred on Channel 2 input.
6
Channel 1 Fault
Fault has occurred on Channel 1 input.
5-4
---
(Not used.)
3
Channel 4 OVL
Overload has occurred on Channel 4 input.
2
Channel 3 OVL
Overload has occurred on Channel 3 input.
1
Channel 2 OVL
Overload has occurred on Channel 2 input.
0
Channel 1 OVL
Overload has occurred on Channel 1 input.
9
8
10
11
12
13
14
15
To OVLR bit in
Operati on Status
Register
OR
1
0
2
3
4
5
6
7
:OVLR?
Overload Event Register
:OVL
:OVL?
Overload Event Enable
(Mask) Register
Chan2
OVL
Chan1
OVL
Chan3
OVL
Chan4
OVL
Chan2
Fault
Chan1
Fault
Chan3
Fault
Chan4
Fault
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...