
:MEASure Commands
20
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
365
:MEASure:TVALue
(see
page 790
)
Query Syntax
:MEASure:TVALue? <value>, [<slope>]<occurrence>[,<source>]
<value> ::= the vertical value that the waveform must cross.
The
value can be volts or a math function value such as dB,
Vs, or V/s.
<slope> ::= direction of the waveform.
A rising slope is indicated
by a plus sign (+).
A falling edge is indicated by a
minus sign (-).
<occurrence> ::= the transition to be reported. If the occurrence
number is one, the first crossing is reported.
If
the number is two, the second crossing is reported,
etc.
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
When the :MEASure:TVALue? query is sent, the displayed signal is
searched for the specified value level and transition. The time interval
between the trigger event and this defined occurrence is returned as the
response to the query.
The specified value can be negative or positive. To specify a negative
value, use a minus sign (- ). The sign of the slope selects a rising (+) or
falling (- ) edge. If no sign is specified for the slope, it is assumed to be
the rising edge.
The magnitude of the occurrence defines the occurrence to be reported.
For example, +3 returns the time for the third time the waveform crosses
the specified value level in the positive direction. Once this value crossing
is found, the oscilloscope reports the time at that crossing in seconds,
with the trigger point (time zero) as the reference.
If the specified crossing cannot be found, the oscilloscope reports
+9.9E+37. This value is returned if the waveform does not cross the
specified value, or if the waveform does not cross the specified value for
the specified number of times in the direction specified.
If the optional source parameter is specified, the current source is
modified.
Return Format
<value><NL>
N O T E
This query is not available if the source is FFT (Fast Fourier Transform).
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...