:FUNCtion Commands
16
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
293
:FUNCtion:SOURce1
(see
page 790
)
Command Syntax
:FUNCtion:SOURce1 <value>
<value> ::= {CHANnel<n> | GOFT}
<n> ::= {1 | 2 | 3 | 4} for 4ch models
<n> ::= {1 | 2} for 2ch models
The :FUNCtion:SOURce1 command is used for any :FUNCtion:OPERation
selection (including the ADD, SUBTract, or MULTiply channel math
operations and the FFT transform). This command selects the first source
for channel math operations or the single source for the transforms.
The GOFT parameter is only available for the FFT function. It lets you
specify, as the function input source, the addition, subtraction, or
multiplication of two channels. When GOFT is used, the g(t) source is
specified by the :FUNCtion:GOFT:OPERation, :FUNCtion:GOFT:SOURce1,
and :FUNCtion:GOFT:SOURce2 commands.
Query Syntax
:FUNCtion:SOURce1?
The :FUNCtion:SOURce1? query returns the current source1 for function
operations.
Return Format
<value><NL>
<value> ::= {CHAN<n> | GOFT}
<n> ::= {1 | 2 | 3 | 4} for 4ch models
<n> ::= {1 | 2} for 2ch models
See Also
•
"Introduction to :FUNCtion Commands"
on page 278
•
":FUNCtion:OPERation"
on page 289
•
":FUNCtion:GOFT:OPERation"
on page 285
•
":FUNCtion:GOFT:SOURce1"
on page 286
•
":FUNCtion:GOFT:SOURce2"
on page 287
N O T E
Another shorthand notation for SOURce1 in this command/query (besides SOUR1) is
SOUR.
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...