:SEARch Commands
26
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
543
:SEARch:SERial:IIC:MODE
(see
page 790
)
Command Syntax
:SEARch:SERial:IIC:MODE <value>
<value> ::= {READ7 | WRITE7 | NACKnowledge | ANACk | R7Data2
| W7Data2 | RESTart | READEprom}
The :SEARch:SERial:IIC:MODE command selects the type of IIC
information to find in the Lister display:
•
READ7 — searches for 7- bit address frames containing
Start:Address7:Read:Ack:Data. The value READ is also accepted for
READ7.
•
WRITe7 — searches for 7- bit address frames containing
Start:Address7:Write:Ack:Data. The value WRITe is also accepted for
WRITe7.
•
NACKnowledge — searches for missing acknowledge events.
•
ANACk — searches for address with no acknowledge events.
•
R7Data2 — searches for 7- bit address frames containing
Start:Address7:Read:Ack:Data:Ack:Data2.
•
W7Data2 — searches for 7- bit address frames containing
Start:Address7:Write:Ack:Data:Ack:Data2.
•
RESTart — searches for another start condition occurring before a stop
condition.
•
READEprom — searches for EEPROM data reads.
When searching for events containing addresses, address values are
specified using the :SEARch:SERial:IIC:PATTern:ADDRess command.
When searching for events containing data, data values are specified using
the :SEARch:SERial:IIC:PATTern:DATA and
:SEARch:SERial:IIC:PATTern:DATA2 commands.
Query Syntax
:SEARch:SERial:IIC:MODE?
The :SEARch:SERial:IIC:MODE? query returns the currently selected mode.
Return Format
<value><NL>
<value> ::= {READ7 | WRITE7 | NACK | ANAC | R7D2 | W7D2 | REST
| READE}
See Also
•
Chapter 26
, “:SEARch Commands,” starting on page 531
N O T E
The short form of READ7 (READ7), READEprom (READE), and WRITe7 (WRIT7) do not
follow the defined Long Form to Short Form Truncation Rules (see
page 792
).
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...