:ACQuire Commands
7
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
185
:ACQuire:COMPlete
(see
page 790
)
Command Syntax
:ACQuire:COMPlete <complete>
<complete> ::= 100; an integer in NR1 format
The :ACQuire:COMPlete command affects the operation of the :DIGitize
command. It specifies the minimum completion criteria for an acquisition.
The parameter determines the percentage of the time buckets that must be
"full" before an acquisition is considered complete. If :ACQuire:TYPE is
NORMal, it needs only one sample per time bucket for that time bucket to
be considered full.
The only legal value for the :COMPlete command is 100. All time buckets
must contain data for the acquisition to be considered complete.
Query Syntax
:ACQuire:COMPlete?
The :ACQuire:COMPlete? query returns the completion criteria (100) for
the currently selected mode.
Return Format
<completion_criteria><NL>
<completion_criteria> ::= 100; an integer in NR1 format
See Also
•
"Introduction to :ACQuire Commands"
on page 183
•
":ACQuire:TYPE"
on page 195
•
":DIGitize"
on page 159
•
":WAVeform:POINts"
on page 640
Example Code
' AQUIRE_COMPLETE - Specifies the minimum completion criteria for
' an acquisition.
The parameter determines the percentage of time
' buckets needed to be "full" before an acquisition is considered
' to be complete.
myScope.WriteString ":ACQuire:COMPlete 100"
See complete example programs at:
Chapter 38
, “Programming Examples,”
starting on page 799
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...