Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
20
:MEASure Commands
Introduction to
:MEASure
Commands
The commands in the MEASure subsystem are used to make parametric
measurements on displayed waveforms.
Measurement Setup
To make a measurement, the portion of the waveform required for that
measurement must be displayed on the oscilloscope screen.
Measurement Error
n/a
:MEASure:VTIMe?
<vtime>[,<source>]
(see
page 374
)
<vtime> ::= displayed time from
trigger in seconds in NR3 format
<source> ::= {CHANnel<n> |
FUNCtion | MATH | WMEMory<r>} for
DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | FUNCtion | MATH |
WMEMory<r>} for MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<return_value> ::= voltage at the
specified time in NR3 format
:MEASure:VTOP
[<source>] (see
page 375
)
:MEASure:VTOP?
[<source>] (see
page 375
)
<source> ::= {CHANnel<n> |
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= voltage at the
top of the waveform in NR3 format
:MEASure:WINDow
<type> (see
page 376
)
:MEASure:WINDow? (see
page 376
)
<type> ::= {MAIN | ZOOM | AUTO}
Table 70 :MEASure Commands Summary (continued)
Command
Query
Options and Query Returns
Measurement Type
Portion of waveform that must be displayed
period, duty cycle, or frequency
at least one complete cycle
pulse width
the entire pulse
rise time
rising edge, top and bottom of pulse
fall time
falling edge, top and bottom of pulse
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...