:ACQuire Commands
7
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
191
:ACQuire:SEGMented:INDex
(see
page 790
)
Command Syntax
:ACQuire:SEGMented:INDex <index>
<index> ::= an integer from 1 to 25 (w/100K memory) in NR1 format
The :ACQuire:SEGMented:INDex command sets the index into the memory
segments that have been acquired.
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command. The number of segments to acquire is set using
the :ACQuire:SEGMented:COUNt command, and data is acquired using the
:DIGitize, :SINGle, or :RUN commands. The number of memory segments
that have been acquired is returned by the :WAVeform:SEGMented:COUNt?
query. The time tag of the currently indexed memory segment is returned
by the :WAVeform:SEGMented:TTAG? query.
The maximum number of segments may be limited by the memory depth
of your oscilloscope. For example, an oscilloscope with 100K memory
allows a maximum of 25 segments.
Query Syntax
:ACQuire:SEGMented:INDex?
The :ACQuire:SEGMented:INDex? query returns the current segmented
memory index setting.
Return Format
<index><NL>
<index> ::= an integer from 1 to 25 (w/100K memory) in NR1 format
See Also
•
":ACQuire:MODE"
on page 187
•
":ACQuire:SEGMented:COUNt"
on page 190
•
":DIGitize"
on page 159
•
":SINGle"
on page 178
•
":RUN"
on page 176
•
":WAVeform:SEGMented:COUNt"
on page 647
•
":WAVeform:SEGMented:TTAG"
on page 648
•
":ACQuire:SEGMented:ANALyze"
on page 189
•
"Introduction to :ACQuire Commands"
on page 183
Example Code
' Segmented memory commands example.
' -------------------------------------------------------------------
N O T E
This command is available when the segmented memory option (Option SGM) is enabled.
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...