:MARKer Commands
19
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
321
:MARKer:X1Y1source
(see
page 790
)
Command Syntax
:MARKer:X1Y1source <source>
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= {1 | 2}
The :MARKer:X1Y1source command sets the source for the cursors. The
channel you specify must be enabled for cursors to be displayed. If the
channel or function is not on, an error message is issued.
If the marker mode is not currently WAVeform (see
":MARKer:MODE"
on
page 319):
•
Sending a :MARKer:X1Y1source command will put the cursors in the
MANual mode.
•
Setting the source for one pair of markers (for example, X1Y1) sets the
source for the other (for example, X2Y2).
If the marker mode is currently WAVeform, the X1Y1 source can be set
separate from the X2Y2 source.
If :MARKer:MODE is set to OFF or MANual, setting :MEASure:SOURce to
CHANnel<n>, FUNCtion, MATH, or WMEMory<r> will also set
:MARKer:X1Y1source and :MARKer:X2Y2source to this value.
Query Syntax
:MARKer:X1Y1source?
The :MARKer:X1Y1source? query returns the current source for the
cursors. If all channels are off or if :MARKer:MODE is set to OFF, the
query returns NONE.
Return Format
<source><NL>
<source> ::= {CHAN<n> | FUNC | WMEM<r> | NONE}
See Also
•
"Introduction to :MARKer Commands"
on page 318
•
":MARKer:MODE"
on page 319
•
":MARKer:X2Y2source"
on page 323
•
":MEASure:SOURce"
on page 361
N O T E
MATH is an alias for FUNCtion. The query will return FUNC if the source is FUNCtion or
MATH.
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...