:DIGital<d> Commands
12
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
247
:DIGital<d>:POSition
(see
page 790
)
Command Syntax
:DIGital<d>:POSition <position>
<d> ::= 0 to (# digital channels - 1) in NR1 format
<position> ::= integer in NR1 format.
The :DIGital<d>:POSition command sets the position of the specified
channel. Note that bottom positions might not be valid depending on
whether digital buses, serial decode waveforms, or the zoomed time base
are displayed.
Query Syntax
:DIGital<d>:POSition?
The :DIGital<d>:POSition? query returns the position of the specified
channel.
If the returned value is "- 1", this indicates there is no space to display the
digital waveform (for example, when all serial lanes, digital buses, and the
zoomed time base are displayed).
Return Format
<position><NL>
<position> ::= integer in NR1 format.
See Also
•
"Introduction to :DIGital<d> Commands"
on page 243
Channel Size
Position
Top
Bottom
Large
0-7
7
0
Medium
0-15
15
0
Small
0-31
31
0
N O T E
This command is only valid for the MSO models.
Summary of Contents for InfiniiVision 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 24: ...24 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide...
Page 38: ...38 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 48: ...48 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 62: ...62 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 206: ...206 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 242: ...242 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 250: ...250 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 270: ...270 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 14 DVM Commands...
Page 312: ...312 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 17 HARDcopy Commands...
Page 332: ...332 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 19 MARKer Commands...
Page 416: ...416 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 22 POD Commands...
Page 424: ...424 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 23 RECall Commands...
Page 446: ...446 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 24 SAVE Commands...
Page 588: ...588 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 28 TIMebase Commands...
Page 692: ...692 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 31 WGEN Commands...
Page 758: ...758 Agilent InfiniiVision 2000 X Series Oscilloscopes Programmer s Guide 34 Error Messages...