DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 30
06-2008 Type
5367/100
Agfa Company Confidential
(13) Leave the “service command session”.
(14) Repeat step “2.5.6 Performing the Afssaps / EUREF Calibration Check.
•
If this time all “Deviation compared to DICOM (%)” values are < 10%, the
“Afssaps / EUREF calibration procedure” is finished.
•
If one (or more) “Deviation compared to DICOM (%)” value(s) are
≥
10%,
repeat steps 2.5.6 and 2.5.7 up to three times. See IMPORTANT note
below.
IMPORTANT:
If after three repetitions one or more “Deviation compared to DICOM (%)” values are
still
≥
10%, start the complete procedure from scratch. Refer to section 2.5.