![AGFA DRYSTAR AXYS Service Manual Download Page 324](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651324.webp)
DD+DIS033.07E
Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
3.2.1
New Feature: Compliance Film Calibration
With software < 1.6.2_C1 the printer does not always pass the acceptance
procedure as defined by Afssaps / EUREF: The film calibration is not necessarily
accurate enough.
The “compliance film calibration” may be required as a last step of an acceptance
procedure that meets the special criteria set by Afssaps / EUREF.
It can be executed via new browser menu “Service Engineer Tools – Enter Service
command session – 6 Calibration – 6 Compliance Film Calibration”.
For details refer to Technical Documentation Drystar Axys, DD+DIS033.07E,
Adjustments and Calibrations, edition 1, revision 1 (and higher revisions).
Service - Calibration
Film calibration
1.
Internal densitometer calibration
2.
Thermal print head calibration
3.
Horizontal alignment
4.
Smart profile calibration
5.
Compliance film calibration
6.
Go back
This page is generated on 21-05-2008 12:33:33
Step Target values Measured values Step Target values Measured values
00
0.16
01
0.33
02
0.49
03
0.66
04
0.83
05
0.99
06
1.16
07
1.33
08
1.50
09
1.66
10
1.83
11
2.00
12
2.16
13
2.33
14
2.50
15
2.66
16
2.83
17
3.00
Service - Calibration
Compliance film calibration
Start calibration
Go back
Reset values
0.16
0.33
This page is generated on 22-05-2008 08:46:20
0.49
0.66
0.83
0.99
1.16
1.33
1.50
1.66
1.83
2.00
2.16
2.33
2.50
2.66
2.83
3.00
5367_SIB04_01.CDR
Figure 2
Edition 1, Revision 3
DRYSTAR AXYS
Chapter 3.8 / 10
03-2011 Type
5367/100
Agfa Company Confidential
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......