2.2.2
Infocounter Analysis via Printers Infocounters Report
Required Tools
•
Service PC, required to connect to the printer and analyze the infocounter.
536632el.cdr
Printer
Service PC
(1) Connect the Service PC to the
network. For details see chapter 3.2,
Machine specific tools
(2) Start the browser (Internet Explorer
≥
5)
(3) Enter the IP address of the printer in
the address window.
(4) Select menu "Service Engineer
Tools - Printer Infocounters Report"
(5) Analyze the infocounter file as
described in section 2.2.4,
How to evaluate the Infocounter
Figure 1
2.2.3
Infocounter Analysis via Service Command Session
Required Tools
•
Service PC, required to connect to the printer and analyze the infocounter
536632el.cdr
Printer
Service PC
(1) Connect the Service PC to the
network
(2) Start the browser (Internet Explorer
≥
5)
(3) Enter the IP address of the printer in
the address window.
(4) Select menu "Service Engineer
Tools - Enter Service Command
Session - 7 Service Actions - 5
Logging - 2 Infocounters"
Figure 2
(5) Save the infocounter file to the service PC or the CF card.
(6) Unzip the infocounters ZIP file
(7) Analyze the infocounter.txt file as described in section 2.2.4,
How to evaluate the Infocounter
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......