DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 12
06-2008 Type
5367/100
Agfa Company Confidential
2.3
Print Head (TH) Calibration
WHEN TO BE
PERFORMED
•
The Print Head Profile calibration is necessary whenever vertical stripes are not
longer acceptable:
Vertical stripes are unacceptable in case they are visible in the medical service
test image.
REQUIRED TOOLS:
•
One of following, calibrated densitometers: McBeth TR-924, Gretag D200-2, X-
Rite 301, X-Rite 310, X-Rite 331, X-Rite 341
•
Lint free cloth
•
Isopropyl - alcohol
INSTRUCTIONS
(1)
Print the Medical service test image.
(2)
Clean the thermal head with a cleaning tissue lightly moistened with Isopropyl -
alcohol. The last cleaning action must be carried out in one way.
(3)
In the webpages - menu 'Service Engineer Tools - Enter Service Command
session' select '6. Calibration - 3. Print Head Calibration'.
(4)
Follow the instructions on the WebPages. Refer to Figure 4.
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......