DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 28 of 72
03-2011 Type
5367/100
Agfa Company Confidential
Cure3
Connect connector P1 on TH
Reason4
Connector P43 on MOD08 baseboard (Devil or Devius) not connected
Cure4
Connect connector P43 on MOD08 baseboard (Devil or Devius)
Reason5
PSU
defect
Cure5
Replace
Part
MOD09
Reason6
OSB
defect
Cure6
Replace
Part
OSB
Reason7
TH
defect
Cure7
Replace
Part
MOD05