Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 41 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC308002
: devilmodule.keypad.connected
Type
Defect
Context
E308002
Action
SeeManual
Reason1
Connector P8 not connected
Cure1
Connect
P8
Reason2
MOD08 baseboard (Devil or Devius) module defect
Cure2
Replace
Part
MOD08
HC308003
: devilmodule.th.connected
Type
Defect
Context
E308003
Action
SeeManual
Reason1
Connector P43 not connected
Cure1
Connect
P43
Reason2
Connector P1 on TH not connected
Cure2
Connect P1 on TH
Reason3
MOD08 baseboard (Devil or Devius) module defect
Cure3
Replace
Part
MOD08
Reason4
TH
defect
Cure4
Replace
Part
MOD05
IMPORTANT:
This fault may have following additional reason and cure:
Reason: Ambient temperature too low when switching on the printer.
Cure: Ensure that ambient temperature is within specifications (+15 ° C to +30 ° C)
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......