![AGFA DRYSTAR AXYS Service Manual Download Page 144](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651144.webp)
DD+DIS033.07E
Repair and Service
Tools and Auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
(2) Select the desired parameter, in this case ‘Perception LUT = Kanamori_like’ and
set the ‘Kanamori_like value’ to 200.
(3) Check the checkbox before ‘Perception LUT’.
(4) Enter AE title and nickname (nickname = the name the stuff uses for this
modality. The nickname will be displayed in the queue). Refer to Figure 36.
(5) Select
‘Submit’
Now whenever the host with AE title ‘CTSCAN200’ sends a print job to the printer, the
taste LUT will be adapted to ‘Kanamori_like 200’.
NOTE:
•
In case the checkbox is checked before a parameter in the DICOM profile editor,
this parameter is used, no matter whether the modality sends the parameter or
not (‘overwrite functionality’)
•
In case the checkbox is not checked, the value defined in the host profile is only
used, in case it is not sent by the modality.
•
Image adaptations should be made preferably at the DICOM modality. According
to DICOM specifications, controlling the printer completely is done at the
modality.
•
The description below about creating a DICOM profile should only be required,
in case the modality does not allow to adapt a certain parameter or does not
provide it.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 3.2 / 38
10-2007 Type
5367/100
Agfa Company Confidential
Summary of Contents for DRYSTAR AXYS