![AGFA DRYSTAR AXYS Service Manual Download Page 192](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651192.webp)
DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 26 of 72
03-2011 Type
5367/100
Agfa Company Confidential
Part
MOD09
Reason7
TH
defect
Cure7
Replace
Part
MOD05
HC305003
: thmodule.elabel.present
Type
Failure
Context
W305003
Action
None
Reason1
TH Elabel not connected
Cure1
Connect TH elabel
Reason2
Elabel
defect
Cure2
Replace
Part
MOD05
HC305004
: thmodule.elabel.config
Type
Failure
Context
W305004
Action
None
Reason
Elabel
defect
Cure
Replace
Part
MOD05