![AGFA DRYSTAR AXYS Service Manual Download Page 225](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651225.webp)
DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 59 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC310001
: displaykeypadmodule.elabel.config
Type
Failure
Context
W310001
Action
None
Reason
Display Elabel defect
Cure
Replace
Part
MOD10
MODULE : inputsensormodule
Elabel : No
HC311000
: inputsensormodule.s4.flagproblem
Type
Failure
Context
W311000
Action
None
Reason
Internal jam, input sensor S4 isn't responding
Cure
Remove
jam
HC311001
: inputsensormodule.s13.flagproblem
Type
Failure
Context
W311001
Action
None
Reason
Internal jam, input sensor S13 isn't responding
Cure
Remove
jam