![AGFA DRYSTAR AXYS Service Manual Download Page 159](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651159.webp)
DD+DIS033.07E
Repair and Service
Tools and Auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Part 1: Recording the DICOM data
(1) Change the IP address of the
printer to any free IP address in
the network (e.g.
192.9.200.202)
(2) Give the service PC the former
IP address of the printer (e.g.
192.9.200.201)
(3) Start
ADVP
(4) Select 'Basic Grayscale' in the
'Start up configuration' screen.
(5) Select 'Application - Comms
Recorder - Start'
(6) In the following window fill in
- Called (Printer) AE Title (this is
the AE title of the printer),
- 'Called (Host) AE Title',
- ADVP Listen port (usually 104:
this is the port the modality
sends DICOM messages),
- Printer IP address (this is the
IP address of the printer)
- printer listen port (usually
104).
(7) Select okay. Now ADVP is
prepared for recording the
communication between
modality and the printer.
Figure 53: ADVP as Comms Recorder
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 3.2 / 53
10-2007 Type
5367/100
Agfa Company Confidential
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......