![AGFA DRYSTAR AXYS Service Manual Download Page 231](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651231.webp)
DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 65 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC314005
: gearboxmodule.s3
Type
Defect
Context
E314005
Action
SeeManual
Reason1
Sensor S3 not connected
Cure1
Connect sensor S3
Reason2
Roller drive module defect
Cure2
Replace
Part
MOD14
HC314006
: gearboxmodule.connected
Type
Defect
Context
E314006
Action
SeeManual
Reason1
Connector XS12 not connected
Cure1
Connect Connector XS12
Reason2
gearboxmodule drive module defect
Cure2
Replace
Part
MOD14
HC314020
: gearboxmodule.working
Type
Defect
Context
14
Action
Replace
Reason
Roller drive module defect
Cure
Replace
Part
MOD14