![AGFA DRYSTAR AXYS Service Manual Download Page 206](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651206.webp)
DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 40 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC307021
: drumdrivemodule.m1.shortcircuit
Type
Defect
Context
E307021
Action
SeeManual
Reason
Drum drive module defect
Cure2
Replace
Part
MOD07
MODULE : devilmodule
Elabel : Yes
HC308000
: devilmodule.dsh.connected
Type
Defect
Context
E308000
Action
SeeManual
Reason1
Connector P9 not connected
Cure1
Connect
P9
Reason2
MOD08 baseboard (Devil or Devius) module defect
Cure2
Replace
Part
MOD08
HC308001
: devilmodule.psu.connected
Type
Defect
Context
E308001
Action
SeeManual
Reason1
Connector P15 not connected
Cure1
Connect
P15
Reason2
MOD08 baseboard (Devil or Devius) module defect
Cure2
Replace
Part
MOD08