![AGFA DRYSTAR AXYS Service Manual Download Page 389](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651389.webp)
DD+DIS032.07E
Maintenance
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 9 / 11
09-2007 Type
5367
Agfa Company Confidential
2.2.4
How to evaluate the Infocounter
Infocounters.txt evaluation
What to check in the
infocounter
Comment
1.1 Serial number and
Installation date
Compare device serial number with chapter 8
‘Manufacturing Standard Modification’ to determine whether
the device possibly requires a modification.
Refer to 3.3 (infocounter.txt), ‘HW Modification History’, to
check already made modifications.
1.4 Software Info
Compare the currently installed SW version with the latest
SW version described in chapter 3.8, SW releases,
patches. In case a recommended SW patch is available,
install this SW patch.
Refer to section 4.3.
Infocounters.txt evaluation
What to check in the
infocounter
Comment
2.3 Reliability
Should normally be not higher than 0.05. Higher fault rates
need a thorough investigation.
3.3 Hardware
Modification History
By comparing the status of the device with the available
modifications (refer to Technical Documentation chapter 8)
the exact hardware status can be determined.
3.4 Software Modification
History
By checking the software modification history it can be
determined, whether a recent software upgrade solved a
problem, that occurred quite often in the error list in the
past
5.3 Retries
Many retries (> 1%, based on relative prints) have to be
investigated: They may reduce the throughput of the
device.
5.7 Error History
Check which errors occurred in the last days, how often
they appeared as well as the CBF (cycles between failures)
of these errors. This gives an overview of the current status
of the machine.
5.8 Error list (sorted by
relative) and
5.9 Error list (sorted by
total):
Troubleshoot these errors with the help of the technical
documentation, chapter 3.3, ‘Troubleshooting’.
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......