![AGFA DRYSTAR AXYS Service Manual Download Page 244](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651244.webp)
DD+DIS033.07E
Repair and Service
Electrical and Mechanical Codes, Fuses, LEDs
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 3.4 / 6
10-2007 Type
5367/100
Agfa Company Confidential
Jumper C4
Setting
Function
1-2
open
ID Bit 3
3-4
open
ID Bit 2
5-6
open
ID Bit 1
7-8
open
ID Bit 0
9-10
open
Not used
11-12
open
Write Protect
Table 4
2 Fuses
The printer has no fuses that can be replaced.
If one of the fuses in the power supply is defective, the whole Power supply unit must
be replaced - refer to section 3.5.
3 LEDs
The following tables show the status of the LED indications and the corresponding
function: