![AGFA DRYSTAR AXYS Service Manual Download Page 57](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651057.webp)
DD+DIS033.07E
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 1 / 21
06-2008 Type
5367/100
Agfa Company Confidential
Enclosed an example where it is required to configure two logical SCP printers:
•
Printing is initiated at a PACS workstation.
•
Two types of modalities as input.
o
Mammography images: Interpolation cubic bell with
smoothfactor 1.1
o
CT images: Interpolation highres cubic with
smoothfactor -0.9.
Problem:
The PACS workstation has one AE title only. The DRYSTAR AXYS does not
recognize, for which type of modality (CT or Mammo) the user wants to print.
Solution (see "
Figure 14: Example for logical printer setup
" next page):
•
Create one logical printer 'DSAXYSMammo' and one 'DSAXYSCT.
•
Create two different host profiles for the two logical printers with the required
settings.
•
In the PACS workstation setup both printers: The user has to select the proper
logical printer.
NOTE:
In the example of Figure 14
X
the checkboxes before the "Kernel" parameters are set.
This setting prevents, that the modality can overrule the adjusted parameter.
It always depends on the application, whether this is desired or not.
This has to be discussed with the modality service engineer and/or the customer.
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......