DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 20
06-2008 Type
5367/100
Agfa Company Confidential
Overview of the Afssaps / EUREF calibration procedure:
Perform densitometer
calibration acc. to densitometer
manufacturer instructions.
Adjust densitometer in 'Key-operator menu -
2 Change settings – 5 Image Quality
- 1 General settings'
Drystar AXYS Mammo calibration procedure for Afssaps/EUREF
YES
YES
YES
No
No
Densitometer
calibration ok?
Printer installation date
= today?
DRC ok?
Deviations < 10%?
Check calibration of densitometer. Details see section 2.5.1.
Check configured default densitometer. Details see section 2.5.2.
Used densitometer
= default densitometer
in the printer?
Reset CDM, TH, CFC calibrations. Details see section 2.5.3
Print TG18 QPC pattern. Measure 18 density values.
Enter values in 'EUREF&AFSSAPS GSDF check.xls'. Details see 2.5.6.
Perform density reproduction check (DRC). Details see section 2.5.5.
5. Film calibration for mammo trays (2 x)
Perform CFC calibration.
Details see section 2.5.7.
Repeat up to 3 times
if required.
Perform following calibrations. Details see section 2.5.4:
1. Film calibration for mammo trays (2 x)
2. CDM calibration
3.
Film calibration for mammo trays (1 x)
4. Print Head calibration (1x)
No
No
No
YES
CFC_calibration_axys.cdr
Finished
YES
Figure 11
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......