DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 27 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC305005
: thmodule.device.current
Type
Defect
Context
E305005
Action
SeeManual
Reason1
Connector P3 on PSU not connected
Cure1
Connect P3 on PSU
Reason2
Connectors P2 and P3 on OSB not connected
Cure2
Connect connectors P2 and P3
Reason3
Connectors P3 on TH not connected
Cure3
Connect connector P3 on TH
Reason4
Connector P2 on TH not connected
Cure4
Connect connector P2 on TH
Reason5
Connector P1 on TH not connected
Cure5
Connect connector P1 on TH
Reason6
Connector P43 on MOD08 baseboard (Devil or Devius) not connected
Cure6
Connect connector P43 on MOD08 baseboard (Devil or Devius)
Reason7
PSU
defect
Cure7
Replace
Part
MOD09
Reason8
OSB
defect
Cure8
Replace
Part
OSB
Reason9
TH
defect
Cure9
Replace
Part
MOD05
HC305006
: thmodule.powerlogic
Type
Defect
Context
E305006
Action
SeeManual
Reason1
Connector P3 on PSU not connected
Cure1
Connect P3 on PSU
Reason2
Connectors P2 on TH not connected
Cure2
Connect connector P2
Reason3
Connector P1 on TH not connected
Summary of Contents for DRYSTAR AXYS