DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 56 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC309004
: powersupplymodule.powermotor.current
Type
Defect
Context
E309004
Action
SeeManual
Reason1
Actuators power cables short circuited
Cure1
Check actuators power cables
Reason2
PSU
module
defect
Cure2
Replace
Part
MOD09
HC309005
: powersupplymodule.temperature.warning
Type
Failure
Context
W309005
Action
None
Reason
PSU temperature high
Cure
Cool down the PSU
HC309020
: powersupplymodule.powerth
Type
Defect
Context
09
Action
Replace
Reason
PSU
module
defect
Cure
Replace
Part
MOD09