![AGFA DRYSTAR AXYS Service Manual Download Page 182](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651182.webp)
DD+DIS033.07E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 16 of 72
03-2011 Type
5367/100
Agfa Company Confidential
HC303014
: upperpickupmodule.connected
Type
Failure
Context
W303014
Action
None
Reason1
XS1
disconnected
Cure
Connect
XS1
Reason
Upper pickup module defect
Cure
Replace
Part
MOD03
HC303015
: upperpickupmodule.working
Type
Failure
Context
W303015
Action
None
Reason1
Upper pickup module defect
Cure1
Replace
Part
MOD03
Reason2
Film wrinkled in upper tray
Cure2
Remove
wrinkled
film
HC303016
: upperpickupmodule.passed.high
Type
Failure
Context
W303016
Action
None
Reason1
Upper pickup past sensor jammed
Cure1
Free upper pickup past sensor
Reason2
Upper pickup past sensor defect
Cure2
Replace
Part
MOD03