![AGFA DRYSTAR AXYS Service Manual Download Page 248](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651248.webp)
DD+DIS033.07E
Repair and Service
Electrical and Mechanical Codes, Fuses, LEDs
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 3.4 / 10
10-2007 Type
5367/100
Agfa Company Confidential
3.4
Front Panel
Led Color
Function
D1 green
12V power present
D2..D25
green / red
See table below
D3 green
5V power present
Color
Light
Status
Receive Jobs ?
Printing possible ?
Constant
Ready
(stand-by)
No
No
Startup
sequence
No
No
Receiving jobs
Yes
Yes
Calculating
Yes
Yes
Printing
Yes
Yes
Calibrating
Yes
No
Installation
No
No
Green
Blinking
Key-operator
Yes/No
Yes/No
Service mode
No
No
Film jam
Yes
No
Blinking
Warning
Yes
Yes
Red
Constant
Error status
No
No