DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 28
06-2008 Type
5367/100
Agfa Company Confidential
2.5.6
Performing the Afssaps / EUREF Calibration Check
(1) Leave the service command session.
(2) Connect the printer network cable to the network outlet at the wall.
(3) Consult the “connectivity release document” for the modality connection to
DRYSTAR AXYS Mammo for the proper modality configuration to print the “TG18-
PQC modified 4” test image.
(4) Print the “TG18-PQC modified 4” test image from the modality.
(5) Measure the 18 density values and enter them in the Afssaps / EUREF
Calibration Check Excel sheet. See Figure 17.
EUREF/AFSSAPS GSDF check for printer
Step 2
: Check, that deviation is < 10% for all values. If 1 (or more) value > = 10% go to S
Step 1
: Enter the measured values he
Laser:
Measured luminance values
DICOM 3.14 standard
testpattern
Luminance
JND
deviation compared
testpattern
Luminance
JND
dL/L
dL/L
TG18-LN12-# p-value
L
j(L)
dL/L
to DICOM (%)
TG18-LN12 p-value
L
j(L)
dL/L
+10%
-10%
Step # on TG18 image Measured OD LuminanceStep # in CFC calibration screen on printer
1
0
1,798105
98,486
1
0
1,80
98,486
1
3,7
1,7981049
17
2
240
3,831783
142,848
0,722
9,0
2
240
4,17
148,511
0,794
0,874
0,715
2
3,15
3,8317831
16
3
480
7,638348
193,970
0,664
4,0
3
480
8,07
198,537
0,638
0,702
0,575
3
2,78
7,6383476
15
4
720
13,07981
241,359
0,525
3,5
4
720
14,11
248,563
0,544
0,599
0,490
4
2,52
13,079807
14
5
960
21,51446
291,079
0,488
1,0
5
960
23,09
298,589
0,483
0,531
0,434
5
2,29
21,514455
13
6
1200
34,27055
342,479
0,457
3,9
6
1200
36,13
348,614
0,440
0,484
0,396
6
2,08
34,270551
12
7
1440
51,35702
390,683
0,399
2,6
7
1440
54,76
398,640
0,410
0,451
0,369
7
1,9
51,357016
11
8
1680
75,48349
439,328
0,380
1,8
8
1680
81,09
448,666
0,388
0,426
0,349
8
1,73
75,483485
10
9
1920
111,1691
490,605
0,382
3,1
9
1920
117,99
498,692
0,371
0,408
0,334
9
1,56
111,16915
9
10
2160
163,9521
544,108
0,384
7,2
10
2160
169,45
548,718
0,358
0,394
0,322
10
1,39
163,95211
8
11
2400
231,176
592,847
0,340
2,3
11
2400
240,90
598,743
0,348
0,383
0,313
11
1,24
231,17597
7
12
2640
333,7055
646,119
0,363
6,5
12
2640
339,85
648,769
0,341
0,375
0,307
12
1,08
333,70551
6
13
2880
470,959
697,023
0,341
1,8
13
2880
476,63
698,795
0,335
0,369
0,302
13
0,93
470,95902
5
14
3120
649,724
745,196
0,319
3,5
14
3120
665,49
748,821
0,331
0,364
0,298
14
0,79
649,72404
4
15
3360
917,3471
797,368
0,342
4,3
15
3360
926,13
798,846
0,328
0,360
0,295
15
0,64
917,34706
3
16
3600
1265,911
846,462
0,319
1,8
16
3600
1285,79
848,872
0,325
0,358
0,293
16
0,5
1265,9111
2
17
3840
1747,063
895,837
0,319
1,3
17
3840
1782,24
898,898
0,324
0,356
0,291
17
0,36
1747,0633
1
18
4080
2467,38
948,924
0,342
6,0
18
4080
2467,38
948,924
0,322
0,355
0,290
18
0,21
2467,38
0
Result contrast response printer
lineair relation between p-value and JN
a: 0,208441
b: 98,4856
Step 3
: Enter these values in CFC* screen on print
Only required if deviation >= 10% (see step 2)
* CFC = Compliance Film Calibration
0,100
1,000
0
1000
2000
3000
4000
5000
p-values
dL/
L
left monitor
DICOM 3.14
+/- 10%
deviation compared
to DICOM (%)
Figure 17
(6) Check column “Deviation compared to DICOM (%)”: All values have to be < 10%.
- If all values are < 10%, the procedure is finished.
- If one (or more) value(s) are
≥
10 %, proceed with next section.
(7) Repeat this step for the other film tray, if the printer has two different DT2 Mammo
film formats loaded (e.g. 8x10” DT2 Mammo and 10x12” DT2 Mammo).
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......