DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 18
06-2008 Type
5367/100
Agfa Company Confidential
2.5
Compliance Film Calibration as part of Afssaps / EUREF Calibration
Procedure
WHEN TO BE
PERFORMED
The compliance film calibration may be required as last step of the Afssaps / EUREF*
calibration procedure for a DRYSTAR AXYS used for printing mammo images.
The Afssaps / EUREF compliance calibration procedure has to be performed in
following cases:
•
Before the Afssaps / EUREF mammo acceptance test.
•
Each 6 months, if the printer is under control of Afssaps / EUREF, and it did not
pass the Afssaps / EUREF calibration check (see section 2.5.6).
Agence française de sécurité sanitaire des produits de santé
European Reference Organisation for Quality Assured Breast Screening and
Diagnostic
Services
PREREQUISITE
1) The compliance film calibration is only
available for DRYSTAR AXYS
software
≥
1.6.2_C1.
2) A “TG18-PQC modified 4” image has
to be available on the modality.
The Test Pattern can be downloaded
from
Figure 10: TG18-PQC modified 4
PURPOSE
Make film calibration this accurate, that it passes the acceptance criteria as defined
by Afssaps / EUREF:
The default film calibration procedure is not necessarily accurate enough.
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......