8-7
IM 704510-01E
Detailed Display Setup
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
Copying the Auto Window Settings to the Multi Window or Single Window Settings
8.
Turn the rotary knob to select Copy to Multi Window or Copy to Single Window,
and then press the SELECT key.
An Alert dialog box opens.
9.
Turn the rotary knob to select OK, and then press the SELECT key.
Settings such as the center, span, marker, and constant T values are copied to the
multi window or single window.
Selecting the Window to Be Displayed
10.
Press the ESC key to close the Window Parameter Setting dialog box.
11.
Press the WindowNum soft key from the Display menu.
12.
Turn the rotary knob to select the window to be displayed.
Explanation
Window Mode: Auto
In this mode, the constant T value is automatically determined according to the input
data and the window size is automatically set. This mode is used when multiple
histograms are present simultaneously and the constant T value is not fixed (CD in CAV
format) or when the constant T value cannot be determined.
Selecting the Modulation Type
Select from the following.
• EFM:
Modulation type used in CDs.
• EFM+: Modulation type used in DVDs. It is also called 8-16 modulation.
• 1-7:
1-7 RLL modulation type. Used in rewritable disks.
The TA720 provides the following windows for each modulation type.
Modulation Type
Window
Window Size
EFM
3T to 11T
9 (When 2T is OFF)
EFM+
3T to 11T, 14T
10 (When 2T is OFF)
1-7
2T to 9T
8 (When 1T is OFF)