6.7
Adjusting the Phase during Time Interval
Measurement
For a functional description, see 2.5.
Procedure
1.
Press the INPUT key to display the Input menu.
2.
Press the Setting soft key to display the Input Setting menu.
3.
Press the ChB soft key.
4.
Press the Phase Adj soft key.
Phase Adj appears only for the measurement functions of A-to-B time interval,
period A & A-to-B time interval, and pulse width A & A-to-B time interval (pulse
width A
→
A-to-B time interval).
5.
Set the phase time using the rotary knob or numeric keys.
Explanation
Selectable Range of Phase Time
Shift the phase of the CH B input signal in the following range.
0 to 10.0 ns (resolution is 0.1 ns)
Note
The phase of the CH B input signal is adjusted so that the X-axis center of the histogram is
positioned at the center of the main window.