8-10
IM 704510-01E
8.4
Changing the Scale Value of the Histogram and
Time Variation Displays
For a functional description, see 2.6.
Procedure
Auto Scaling
Press the Auto SCALE key.
Note
Auto scaling cannot be performed in the following cases.
•
When auto window is specified in hardware histogram mode.
•
When multi window or auto window is specified in time stamp mode.
•
When in inter-symbol interference analysis mode.
Manual Scaling
1.
Press the SCALE key to display the Scale menu.
Selecting the Scaling Item (only for dual measurement function)
2.
Press the Link soft key to display the Link menu.
3.
Press the Meas1, Meas2, or Link soft key to select the item to be scaled.
Setting the X-Axis Center
4.
Press the X Center soft key and then set the center value using the rotary knob
or numeric keys.
Selecting the X-Axis Span
5.
Press the X Span soft key, and then turn the rotary knob to select the span
value.
Selecting the Y-Axis Maximum
6.
Press the Y High soft key, and then turn the rotary knob to select the Y-axis
maximum.
Selecting the Y-Axis Scale Type
7.
Press the Y Axis soft key to select Lin or Log.