5-11
IM 704510-01E
Setting the Sampling Mode and Measurement Function
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
• A
A
↑
B
↓
:
Measures from the rising edge to the next rising edge and from the
falling edge to the next falling edge of the CH A input signal. At the
same time, measures from the rising edge of the CH A input signal to
the first falling edge of the CH B input signal.
• A
A
↓
B
↓
:
Measures from the rising edge to the next rising edge and from the
falling edge to the next falling edge of the CH A input signal. At the
same time, measures from the falling edge of the CH A input signal to
the first falling edge of the CH B input signal.
Measurement Range of the Pulse Width A & A-to-B Time Interval Measurement
Varies depending on the sampling mode.
• When in time stamp mode
• Pulse width A measurement:
6 ns to 20 ms
• A-to-B time interval measurement: 0 ns to 20 ms
• When in hardware histogram mode
• Pulse width A measurement:
6 ns to 3.2
µ
s
• A-to-B time interval measurement: 0 ns to 3.2
µ
s
• When in inter-symbol interference analysis mode
• Pulse width A measurement:
10 ns to 3.2
µ
s
• A-to-B time interval measurement: 0 ns to 3.2
µ
s
Continuous Measurement Conditions
When the measurement edge interval of the CH A input signal is 20 ns or more and the
measurement edge of CH B to the measurement edge of the CH A is 13 ns or more,
measurements are made continuously.
CH A
CH B
CH A
CH B
CH A
CH B
20 ns or less
Continuous
measurement
Non-continuous
measurement
20 ns or more
When set to A A
B
↑
Measured
When A-to-B
time interval
measurement
is not continuous
13 ns or less
20 ns or more
20 ns or more
Measured
Measured
Measured
Measured
Measured
Measured
13 ns or more
Measured
Measured
Measured
Measured
Measured
Measured Measured
Measured
Note
•
When the polarity/slope is set to both polarities, histograms and statistical values can be
displayed separately for the measured values of each polarity/slope.
•
When set to A
A B
Separates A
A
↑
B
↑
, A
A
↓
B
↑
, and A
A B
•
When set to A
A B
Separates A
A
↑
B
↓
, A
A
↓
B
↓
, and A
A B
•
Changing the measurement function clears the data that has been acquired to the memory up
to that point.
5.7 Setting the Pulse Width A & A-to-B Time Interval Measurement