6-10
IM 704510-01E
6.5
Setting the Block Sampling
For a functional description, see 2.5.
Procedure
1.
Press the SAMPLE key to display the Sample menu.
Block does not appear when the sampling mode is set to inter-symbol interference
analysis mode (because block sampling cannot be performed).
2.
Press the Block soft key to display the Block Sample selection menu.
3.
Press the Mode soft key to select ON.
Setting the Number of Measurement Blocks
4.
Set the number of measurement blocks using the rotary knob or numeric keys.
Setting the Rest Time
5.
Press the RestMode soft key to display the RestMode selection menu.
6.
Press the appropriate soft key to select the rest mode.
• When Rest Mode Is Set to Time
7.
Press the RestTime soft key.
8.
Set the rest time using the rotary knob or numeric keys.
• When Rest Mode Is Set to Event
7.
Press the RestEvent soft key.
8.
Set the event size to rest using the rotary knob or numeric keys.
Explanation
Selectable Range of Number of Blocks
• When in time stamp mode
The selectable range varies depending on the arming source and rest mode as
follows.
Auto
Ext
OFF
Time
Event
2 to 1000 2 to 250
2 to 250
2 to 250
–
–
Rest mode
Arming source
• When in hardware histogram mode: 2 to 1000